Introduction
Glucose-6-phosphate dehydrogenase (G6PD) is an essential enzyme that protects red blood cells from oxidative damage caused by certain drugs, diseases, and foods.1,2 The X-linked human G6PD gene is highly polymorphic with many mutations resulting in reduced enzyme activity in red blood cells or G6PD deficiency. Exposure to oxidative agents can induce hemolysis in red blood cells with low G6PD activity levels and cause severe anemia, sometimes requiring blood transfusion or causing irreversible renal damage and even mortality, if not managed promptly. Glucose-6-phosphate dehydrogenase deficiency presents clinically in the neonate as jaundice resulting from hyperbilirubinemia; this may lead to kernicterus, a form of brain damage.3,4 Several medications including rasburicase- and 8-aminoquinoline–based antimalarial drugs, such as primaquine, are known to cause clinically significant hemolysis in G6PD-deficient individuals. A high of dose primaquine (a 7-or 14-day regimen) is required to cure patients of Plasmodium vivax malaria. If a patient is not cured of P. vivax, they are at risk of relapse with increasing risk of morbidity and further transmission of the parasite.5,6 Relapse contributes to more than 50% of the disease burden in a community.7,8 A single dose of tafenoquine, another 8-aminoquinoline, when given with chloroquine is capable of curing a patient of P. vivax.9 Tafenoquine, recently approved by the United States Food and Drug Administration (FDA) for radical cure of P. vivax, under the Krintafel label, is indicated for individuals with greater than 70% G6PD activity. Tafenoquine has also been approved by the FDA with a different dosage for prophylaxis, under the Arakoda label, with similar indications for G6PD deficiency.
Many countries will struggle to meet their target malaria elimination goals without broader safe access to radical cure of P. vivax. Diagnostic tests that determine a patient’s G6PD status are needed at or near where they seek treatment.10 Glucose-6-phosphate dehydrogenase deficiency is determined by measuring G6PD activity, adjusted for temperature, in blood normalized for either red blood cell count or hemoglobin. Quantitative testing for G6PD is performed in reference or specialized laboratories using a complex assay on a temperature-regulated instrument because of the large temperature impact on enzyme activity. The most commonly used test for clinical screening is the qualitative fluorescent spot test, which accurately discriminates hemizygous-deficient males and homozygous- or heterozygous-deficient females, who typically have G6PD activity less than 30% of normal. Although this is adequate for males who are either deficient (< 30% normal activity) or normal (> 80% activity), it is inadequate to classify females who can be G6PD deficient, intermediate (30–80% activity), or normal.11–15
Quantitative point-of-care G6PD tests that can be used in low-resource settings can impact health on many levels, including reductions in neonatal morbidity and mortality. It can also impact health by providing access to safe, radical, curative treatment for P. vivax malaria, which can then prevent relapse burden, onward malaria transmission, and accelerate malaria elimination.7–10,16 Both qualitative and quantitative G6PD deficiency tests that can be performed closer to the patients are beginning to emerge.17–19
In 2017, the production of a commonly used reference assay for evaluating new G6PD products, the G6PDH quantitative test by Trinity Biotech, was suspended.20 This study presents the results of a bridging evaluation of the Trinity Biotech kit against a commercially available U.S. Food and Drug Administration–cleared reference assay by Pointe Scientific. The innovative, point-of-care STANDARD G6PD test by SD Biosensor was also evaluated against the same reference assay.
Materials and methods
U.S.-based study.
U.S. donor blood specimens were obtained from Bioreclamation, Inc. (New York, NY), which has donor centers in Miami and New York. Specimens, both venous and capillary, were collected between July 2017 and January 2018 from volunteers of African American origin, at least aged 18 years, and who had signed consent under protocol 2010-017 (Western Institutional Review Board, Washington). Study participants were recruited consecutively and with no knowledge of G6PD status. Participants of African American origin were recruited because of the high prevalence of G6PD deficiency in this population. The sample size was calculated to capture a minimum of 20 G6PD-deficient individuals based on previous studies in a similar population.13 Specimens of 5 mL volume were collected in dipotassium ethylenediaminetetraacetic acid (K2EDTA)−coated anticoagulant venipuncture vacuum tubes (BD Vacutainer®; Becton, Dickinson, Franklin Lakes, NJ). Capillary specimens, 0.2 mL were collected in K2EDTA vacutainers (RAM Scientific, Nashville, TN). All specimens were transported on cold packs and stored at 4°C until tested. Specimen processing occurred within 2 and 4 days of blood collection. Tests for a given comparison were conducted on the same day. All G6PD assays were performed independently and blinded to the G6PD status. A flow chart describing the sampling and testing performed is provided as Supplemental File 1.
Thailand study.
An additional study was conducted by Shoklo Malaria Research Unit (SMRU) in Mae Sot, Thailand, involving samples obtained from two clinical sites located north and south to Mae Sot that serve a migrant population composed of Burman and Karen ethnic groups living along the Myanmar border (Tak Province, northwestern Thailand). Adults with known G6PD status were invited to join the study to achieve a convenience sample of approximately 50 G6PD-deficient volunteers (male and female), 50 G6PD-heterozygous female volunteers, and 50 G6PD-normal volunteers.21 A total of 0.5 mL of venous blood was drawn from the arm, transferred to K2EDTA tubes, and inverted 10 times; samples were kept at 4°C until analysis. Ethical approval was obtained from the ethics committee of the Mahidol University Faculty of Tropical Medicine (FTMEC MO/15/259) and the University of Oxford Tropical Research Ethics Committee (OXTREC 563-15). Written informed consent was obtained from all participants. The blood collection was performed at the clinics.
G6PD activity by reference assays.
Trinity Biotech quantitative G6PD assay.
Glucose-6-phosphate dehydrogenase activity assays were run in duplicate using the quantitative G6PD kit from Trinity Biotech (Cat. No. 345-B) according to the manufacturer’s instructions.
Pointe Scientific quantitative G6PD assay.
Glucose-6-phosphate dehydrogenase activity assays were run in duplicate using the quantitative G6PD kit from Pointe Scientific (Cat. No. G7583) according to the manufacturer’s instructions. Normal, intermediate, and deficient controls from Analytical Control Systems, Inc. (Fishers, IN; Cat. Nos. HC-108, HC-108IN, and HC-108DE, respectively) were run using the same method on each day of testing. The G6PD values were corrected using the temperature correction factor (1.37) at 30°C provided by Pointe Scientific.
For both reference assays, enzyme activity was determined using the UV-1800 temperature-regulated spectrophotometer, set at 30°C, by measuring the change in absorbance rate at 340 nm over 5 minutes. Glucose-6-phosphate dehydrogenase activity values were calculated in U/g hemoglobin (Hb).
Hemoglobin measurement.
Hemoglobin concentration was determined using the HemoCue® Hb 201+ System (No. 121721, Cat. No. 22-601-007; Fisher Scientific, Waltham, MA) for the U.S. specimens or by automated hematology analyzer (CelltacF MEK-8222J/K; Nihon Kohden, Irvine, CA) for the Thailand specimens.
Glucose-6-phosphate dehydrogenase activity by the SD Biosensor STANDARD G6PD test.
The SD Biosensor STANDARD G6PD test is designed to measure the quantitative concentration of total hemoglobin (g/dL) and G6PD enzymatic activity (U/g Hb) in fresh human whole blood based on reflectometry assays. The STANDARD G6PD test consists of two components: a G6PD analyzer instrument and a G6PD test kit (Supplemental File 2). The test kit contains the G6PD test devices (strips), individual extraction buffer tubes, SD Ezi tube+ sample collectors, a single lot-specific code chip, and an instruction manual. Glucose-6-phosphate dehydrogenase activity normalized by hemoglobin was measured with the SD Biosensor STANDARD G6PD test, as per the kit instructions.
Briefly, a device/strip is inserted into the analyzer. Next, using the SD Ezi tube+ sample collector, 10 µL of capillary or venous K2EDTA specimen is transferred to the extraction buffer tube, which is followed by thorough mixing with the SD Ezi tube+ sample collector. A new SD Ezi tube+ sample collector is used to draw a 10 µL mixed specimen and extraction buffer to apply to the test device. After 2 minutes, the analyzer displays the hemoglobin measurement in g/dL and the quantitative ratio of G6PD to hemoglobin (U/g Hb) simultaneously. The STANDARD G6PD test is designed to require only 10 μL of blood, function over the operating temperature range of 15–40°C, and has a dynamic range for specific G6PD activity of 0–20 U/g Hb. The device used for this study has a total hemoglobin range of 7–25 g/dL, more recent versions have a 4–25 g/dL dynamic range claim.
SD Biosensor also provides unitized high and low G6PD controls to support quality assurance of the STANDARD G6PD test. The control reagents are formulated as tablets, which can be applied to the lysis buffer and then run as a normal blood sample. Both levels of G6PD quality control material specific to the STANDARD G6PD test were evaluated. Level 1 control ranges 0–3 U/g Hb G6PD and 4.0–12.0 g/dL of Hb. Level 2 control ranges 6–10 U/g Hb G6PD and 13.0–17.0 g/dL of Hb.
For the data presented here, all blood transfers and dilutions were performed exclusively with disposable SD Ezi tube+ sample collectors as per the instructions and would be used in typical clinical settings. All STANDARD G6PD tests were performed blinded to the reference assays results.
Statistical methods.
Correlation between the reference assay and the investigational device assay was measured by linear regression and the bias was analyzed using Bland–Altman plotting. The clinical performance of the SD Biosensor STANDARD G6PD assay against the Pointe Scientific G6PD quantitative spectrophotometric assay was determined by calculating sensitivity and specificity. Sensitivity and specificity of the STANDARD G6PD assay were calculated as described previously.20 In summary, an adjusted male median was calculated for the Pointe Scientific spectrophotometric gold standard test and defined as 100% activity from which the threshold activities were determined. Area under the curve (AUC) of the receiving operating characteristic (ROC) curve was calculated at different activity thresholds to analyze clinical performance of the STANDARD G6PD test.22 Youden’s index was calculated to determine the optimal cutoff point at various thresholds.22 All statistical analyses were conducted in Stata 13.0 (Statacorp, College Station, TX).
Results
Performance data are presented for three G6PD activity thresholds: 1) < 30% of normal, classified as G6PD deficient, where all hemizygous G6PD-deficient males and all homozygous G6PD-deficient females lie, 2) < 70% of normal, the threshold established in the tafenoquine clinical trials for eligibility for receiving tafenoquine, and 3) < 80% of normal, the current World Health Organization threshold for defining intermediates (30–80% of normal activity) (Table 1). The Pointe Scientific assay was used as the gold standard and the 2 × 2 tables from which sensitivity and specificity values cited in Table 1 are provided as Supplemental Files 3 and 4.
Summary data for the performance of the Trinity Biotech quantitative G6PD assay and the SD Biosensor STANDARD G6PD test compared with the Pointe Scientific quantitative G6PD assay
G6PD diagnostic test | Trinity Biotech quantitative reagent kit | SD Biosensor STANDARD G6PD test | ||
---|---|---|---|---|
Study description | Fresh venous blood (K2EDTA), PATH, USA | Fresh venous blood (K2EDTA), PATH, USA | Fresh venous and capillary blood (both K2EDTA), PATH, USA | Frozen venous blood (K2EDTA), SMRU, Thailand |
No clinical sample (male/female) | 183 | 210 | 100 | 150 |
Male: 128 | Male: 137 | Male: 60 | Male: 42 | |
Female: 55 | Female: 73 | Female: 40 | Female: 108 | |
No deficients | 21 | 25 | 10 | 54 |
No intermediates | 7 | 13 | 8 | 53 |
Adjusted PS median (U/g Hb) | 9.6 | 8.89 | 8.97 | 6.84 |
Normal* PS (SD) (U/g Hb) | 10.4 (2.4) | 9.7 (2.1) | 9.3 (1.8) | 6.7 (1.1) |
Normal* device (SD) (U/g Hb) | 11.4 (3.0)‡ | 10.9 (2.9)† | 11.4 (2.4)† | 6.7 (1.2)† |
Optimal threshold (U/g Hb) equivalent to 30% normal | 2.9 | 2.7 | 2.7 | 2.1 |
% Sensitivity (95% CI) | 100.0 (82.4–100.0) | 100.0 (95.7–100.0) | 100.0 (83.2–100.0) | 100.0 (93.4–100.0) |
% Specificity (95% CI) | 98.2 (94.7–99.6) | 97.0 (94.5–98.5) | 100 (98.0–100.0) | 94.8 (88.3–98.3) |
AUC | > 0.99 | > 0.99 | 1 | 0.99 |
Optimal threshold (U/g Hb) equivalent to 70% normal | 6.7 | 6.2 | 6.3 | 4.8 |
% Sensitivity (95% CI) | 100.0 (86.8–100.0) | 95.5 (89.7–98.5) | 97.2 (85.5–99.9) | 95.0 (88.8–98.4) |
% Specificity (95% CI) | 100.0 (97.7–100.0) | 97.0 (94.5–98.6) | 80.5 (73.6–86.3) | 81.6 (68.0–91.2) |
AUC | 1 | > 0.99 | 0.98 | 0.97 |
Optimal threshold (U/g Hb) equivalent to 80% normal | 7.7 | 7.1 | 7.2 | 5.5 |
% Sensitivity (95% CI) | 89.7 (75.8–97.1) | 95.0 (89.5–98.2) | 97.8 (88.5–99.9) | 96.3 (90.7–99.0) |
% Specificity (95% CI) | 95.1 (90.2–98.0) | 86.3 (81.9–90.1) | 63.6 (55.5–71.2) | 74.4 (58.8–86.5) |
AUC | 0.96 | 0.98 | 0.95 | 0.97 |
AUC = area under the curve; CI = confidence interval; G6PD = glucose-6-phosphate dehydrogenase; K2EDTA = dipotassium ethylenediaminetetraacetic acid; PS = Pointe Scientific; SD = standard deviation; SMRU = Shoklo Malaria Research Unit. Confidence intervals are provided in brackets under each sensitivity and specificity value.
Normal by PS.
Normal by SD Biosensor STANDARD G6PD test.
Estimates for the Trinity Biotech assay.
Correlation between the two quantitative laboratory tests for G6PD activity.
The correlation between the Trinity Biotech quantitative G6PD assay and the Pointe Scientific assay was evaluated with 183 clinical specimens, including 21 G6PD-deficient specimens and seven G6PD-intermediate specimens. The sample size was constrained by the availability of the Trinity quantitative assay product that had been discontinued. Both assays were run at 30°C. The same hemoglobin value measured by HemoCue was used to normalize the G6PD activity values measured by both assays. 30°C Trinity G6PD values were correlated to 37°C Pointe Scientific G6PD values. Linear regression gave a squared correlation coefficient (R2) value of 0.76 and Bland–Altman showed no obvious bias with most specimens falling within an allowable bias of ± 2 U/g Hb for G6PD (Figure 1). Sensitivity and specificity calculated by performing a ROC analysis to identify the optimal values for the 30%, 70%, and 80% thresholds (Table 1, Figure 1B) showed very good concordance in classifying deficient, intermediate, and normal activity between the two assays.
Analytical performance of the SD Biosensor STANDARD G6PD test: temperature and humidity stress testing.
The STANDARD G6PD test consists of an analyzer and an assay kit (Supplemental File 2). After running a blood specimen, the analyzer displays hemoglobin concentration (in g/dL) and temperature-corrected G6PD activity normalized for hemoglobin (U/g Hb at 37°C). Two hundred and ten venous blood specimens in K2EDTA, including 25 G6PD-deficient specimens and 13 intermediate specimens, were used to verify the ability of the product to correct for temperature. All 210 specimens were run on the benchtop in the laboratory at room temperature, approximately 22°C, and a random subset of 100 samples, including 15 G6PD-deficient and five intermediate, were run in a temperature and humidity controlled chamber at 32°C 50% humidity (Supplemental File 5). No statistical difference was observed between results generated at 22°C and 32°C, and the R2 value of the combined data sets was 0.81. Two test results (out of 310) corresponding to specimens identified as normal by the reference assay were misclassified as deficient by the STANDARD G6PD test. This misclassification was not observed in repeat testing, and the root cause was not identified. In addition, 10 of the deficient specimens, five intermediates, and 15 normal samples were run at 37°C, 50% humidity and 37°C, 75% humidity to further stress test the STANDARD G6PD test. An analysis combining all test results, including data measured over the full temperature range of 22–37°C, as individual data points and analyzed for correlation and performance against the reference assay (Figure 2) showed good correlation with an R2 value of 0.85. Hemoglobin measurement on the STANDARD G6PD test showed a tight correlation to the HemoCue, within 1 g/dL Hb and at R2 of 0.87 across all conditions (Figure 2).
Sensitivity and specificity values, based on optimal thresholds determined by ROC curve analysis and the AUC, showed that the STANDARD test performed equivalently to the Trinity assay when evaluated against the Pointe Scientific assay, despite being stressed over broad environmental operating conditions (Table 1).
Performance of the SD Biosensor STANDARD G6PD test in matched venous and capillary specimens.
For 100 of the 210 venous specimens, capillary finger stick specimens were also collected in K2EDTA vacutainer tubes to test the performance of the STANDARD G6PD test compared with the reference Pointe Scientific G6PD assay normalized by the HemoCue hemoglobin measurement for the venous samples. These specimens included 10 G6PD-deficient and eight intermediate samples. The Pointe Scientific assay only has performance claims for venous specimens, so G6PD values for both venous and capillary samples were compared with the G6PD values of Pointe Scientific venous samples. The hemoglobin-normalized G6PD values from both venous and capillary samples showed good correlation against the Pointe Scientific values with an overall R2 of 0.85. No bias was observed for either the venous or the capillary specimens (Supplemental File 6).
Venous hemoglobin values from the STANDARD G6PD test were compared with venous hemoglobin values from the HemoCue device and STANDARD G6PD capillary hemoglobin values were compared with HemoCue capillary values. The correlation for the venous hemoglobin measurements was good between the assays, with an R2 of 0.87. Correlation in hemoglobin values for the capillary samples was initially poor. After 58 samples were run, additional rigorous mixing of the blood samples was introduced before dispensing the aliquot for the assay. Improved correlation was observed for the 42 samples with an R2 value of 0.64 (Supplemental File 6). The high variability in hemoglobin measurement was not reflected in the hemoglobin-normalized G6PD values for the capillary samples. Good sensitivity was attainable at all thresholds, albeit with a drop in specificity for intermediates, whereby normals by Pointe Scientific were classified as intermediates by the STANDARD test (Table 1).
Performance for the STANDARD G6PD test on samples from the Thai–Myanmar border population.
One hundred and fifty frozen, anticoagulated, whole blood, venous samples collected in a previous study were used to evaluate the performance of the STANDARD G6PD test with a geographically distinct population to the previous studies, and in an independent laboratory. The specimens included 54 G6PD-deficient samples (< 30% activity) and 53 with intermediate activity, between 30% and 80% of normal. A good correlation was observed for both normalized G6PD activity and hemoglobin against their respective reference assays, with R2 values of 0.92 and 0.75, respectively. All G6PD values with the exception of two were within 2 U/g Hb variance of the reference value (Figure 3). Good sensitivity and specificity values were observed for deficients with a slight decrease in specificity for intermediates (Table 1). One of 49 female samples with intermediate G6PD activity levels as determined by the reference assay was misclassified as normal by the STANDARD G6PD test and exhibited greater than 2 U/g Hb variance in G6PD values from the reference assay. This sample had reference assay values of 4.6 U/g Hb and 8.0 U/g Hb on the STANDARD G6PD test. Other misclassifications of females with intermediate activity to normal were all within 2 U/g Hb of the reference value.
Performance of SD Biosensor unitized G6PD control reagents.
SD Biosensor also provides unitized high and low G6PD controls to support quality assurance of the STANDARD G6PD test. The performance of the quality control reagents provided by SD Biosensor was also evaluated across two devices and under two temperature and humidity conditions. The reproducibility over five replicates gave a coefficient of variance of under 6.8% in all measurements except for the low control G6PD values, where the variance was higher because of the small values (Table 2).
Performance of quality control reagents provided by SD Biosensor
Low control | High control | |||||||
---|---|---|---|---|---|---|---|---|
Device 1 | Device 2 | Device 1 | Device 2 | |||||
Mean | CV (%) | Mean | CV (%) | Mean | CV (%) | Mean | CV (%) | |
(SD) | (SD) | (SD) | (SD) | |||||
Range | Range | Range | Range | |||||
Room temperature | ||||||||
Hb (g/dL) | 11.2 | 5.2 | 11.3 | 6.5 | 15.6 | 3.4 | 14.8 | 5.2 |
(0.58) | (0.73) | (0.53) | (0.77) | |||||
10.4–12.3 | 10.1–12.5 | 14.7–16.6 | 13.5–15.6 | |||||
G6PD (U/g Hb) | 0.5 | 17.7 | 0.4 | 15.7 | 7.9 | 3.5 | 8.2 | 6.8 |
(0.1) | (0.07) | (0.3) | (0.6) | |||||
0.4–0.7 | 0–0.5 | 7.4–8.4 | 6.9–9.1 | |||||
35°C, 50% Humidity | ||||||||
Hb (g/dL) | 11.3 | 3.4 | 10.9 | 3.1 | 14.6 | 2.9 | 15.2 | 3.4 |
(0.38) | (0.34) | (0.42) | (0.52) | |||||
10.6–11.9 | 10.3–11.3 | 14.0–15.3 | 14.4–16.1 | |||||
G6PD (U/g Hb) | 0.6 | 20.8 | 0.5 | 56.3 | 8.8 | 3.2 | 8.5 | 3.8 |
(0.1) | (0.3) | (0.3) | (0.3) | |||||
0.4–0.8 | 0–0.8 | 8.5–9.2 | 7.9–9.0 |
CV = coefficient of variation; G6PD = glucose-6-phosphate dehydrogenase; SD = standard deviation. Five replicates were run for each condition on two SD Biosensor analyzers.
Discussion
G6PD deficiency is the most common human genetic condition, with potential lifelong clinical implications.1,2,23 Glucose-6-phosphate dehydrogenase deficiency is highly prevalent in malaria-endemic populations, where resources are limited.24 Primaquine and tafenoquine, both 8-aminoquinolines, when given with blood-stage antimalarials can cure patients of P. vivax but may also cause clinically relevant hemolysis in patients with decreased G6PD activity. Ensuring safe, radical curative treatment requires the availability of G6PD deficiency rapid diagnostic tests that are easy-to-use and affordable.10 Qualitative tests developed in recent years attempt to address these needs but they do not accurately identify females heterozygous for G6PD deficiency with intermediate G6PD activity.12,15,25 Some of these females may be at risk of hemolysis, if exposed to the doses of 8-aminoquinolines required to eliminate hypnozoites.16,26 A point-of-care test that can measure G6PD activity normalized by hemoglobin concentration, with a broad operating temperature range could address this limitation. As per the World Health Organization prequalification technical specifications for in vitro diagnostics to measure G6PD activity, such a test must demonstrate the ability to discriminate from normal, G6PD-deficient (< 30%), and G6PD-intermediate (30–80% of normal) activity.27 The performance of the test at the inclusion threshold (> 70%) used in the clinical trials for tafenoquine is also relevant, as this will be used to inform treatment with this drug.9
In this study, two tests for G6PD deficiency were evaluated against the gold standard laboratory quantitative assay for G6PD activity (Pointe Scientific) combined with the gold standard hemoglobin assay (HemoCue). One evaluation compared the Pointe Scientific quantitative assay against the Trinity Biotech G6PDH Quantitative kit, a laboratory-based quantitative assay for G6PD activity that was widely used as a reference gold standard for the evaluation of other G6PD products until production was suspended in 2017. Both the Trinity and Point Scientific assays require temperature-regulated spectrophotometers, a separate hemoglobin measurement, and then manual calculation of G6PD activity normalized for hemoglobin. In the other evaluation, the SD Biosensor STANDARD G6PD point-of-care test was compared with the Pointe Scientific assay. The STANDARD G6PD test concurrently measures G6PD activity and hemoglobin, functions over a broad operating temperature, and provides G6PD activity in international units within 2 minutes without any user calculations required.
The G6PD activity values in the two reference assays were only equivalent when comparing the G6PD activity at 30°C on the Trinity assay with those corrected to 37°C on the Pointe Scientific assay. Both assays used the same hemoglobin measurement. Good sensitivity and specificity were observed for discriminating G6PD-deficient specimens (< 30% normal activity) and G6PD-intermediate specimens (either < 70% or < 80%) using the Pointe Scientific assay as the gold standard, despite an R2 value of 0.76.
The STANDARD G6PD test demonstrated equally good performance with a panel of 210 whole-blood venous specimens from African American volunteers under multiple temperature and humidity operating conditions, ranging from 22°C to 37°C and up to 75% humidity. The capillary G6PD values (U/g Hb) correlated well with the gold standard−matched venous G6PD values for all specimens, despite poor correlation between the two hemoglobin measurements performed on capillary specimens. This indicates that hemoglobin was measured accurately but that deviation from the HemoCue values was a result of genuine differences in the amount of hematocrit drawn into the assay. It is apparent that the measurement of hemoglobin from the same blood sample on which the G6PD assay is performed contributes to the accuracy of the normalized G6PD value (U/g Hb). The STANDARD G6PD test also performed well when evaluated in Thailand with a diverse sample set in a second laboratory, different G6PD genotypes, and good coverage across the full clinically relevant G6PD activity range. For this sample set, the normal G6PD activity was low, both on the gold standard assay and the STANDARD G6PD test, compared with the U.S. sample sets, most likely resulting from the freeze–thaw process; however, this did not affect the concordance between the two tests. The data presented here also has limitations with respect to establishing the robustness of this promising product. Further clinical studies are required in near-patient or intended settings, with larger samples sizes for fresh non-anticoagulated capillary specimens. Future studies should also evaluate the usability of the product by the intended users (not the expert laboratory staff used for this study) and their ability to accurately interpret and record the test results.
The results raise several considerations when assessing new tests for G6PD deficiency. A perfect correlation between two quantitative tests for enzyme activity is very challenging to achieve.13 In addition, there is significant interlaboratory variability in running the same reference assay, which further impacts the ability to establish a perfect correlation between the two assays.28–30 To minimize interlaboratory variability, PATH established a proficiency panel composed of frozen samples. A critical question to resolve is whether robust universal G6PD activity thresholds (in U/g Hb) can be established to discriminate G6PD-deficient (< 30% normal) and intermediate (< 70% or < 80% of normal) activity, or whether it will be necessary to determine these in each setting. This will require more data from multiple settings and more clinical specimens than described here.
Conclusion
Although automated and near-patient quantitative tests for G6PD deficiency have been previously described, many of which rely on fluorescence, none are widely available.18,31 The SD Biosensor STANDARD G6PD test represents a true near-patient product that provides quantitative G6PD results, bringing the possibility of managing all patients—males, females, and neonates—equally with respect to G6PD deficiency, minimizing risk of hemolysis, and improving health outcomes. Specifically, it represents a true opportunity to greatly increase access to radical cure of P. vivax malaria, and thus support malaria elimination, and provides an opportunity to address a gender gap in our understanding of G6PD epidemiology and its clinical implications.
Acknowledgments:
We would like to thank Athena Anderle for her assistance with the manuscript. This work was funded by the U.K. Department for International Development (DFID), grant number 204139, and the Bill & Melinda Gates Foundation, grant number OPP1107113. The findings and conclusions contained within are those of the authors and do not necessarily reflect positions of the Bill & Melinda Gates Foundation or DFID. G. B., G. G., P. P., C. S. C., and F. N. work at the Shoklo Malaria Research Unit, part of the Mahidol Oxford University Research Unit supported by the Wellcome Trust Mahidol Major Overseas Programme–Thailand Unit (grant number 106698).
REFERENCES
- 1.↑
Luzzatto L, Nannelli C, Notaro R, 2016. Glucose-6-phosphate dehydrogenase deficiency. Hematol Oncol Clin North Am 30: 373–393.
- 2.↑
Minucci A, Moradkhani K, Hwang MJ, Zuppi C, Giardina B, Capoluongo E, 2012. Glucose-6-phosphate dehydrogenase (G6PD) mutations database: review of the “old” and update of the new mutations. Blood Cells Mol Dis 48: 154–165.
- 3.↑
Cunningham AD, Hwang S, Mochly-Rosen D, 2016. Glucose-6-phosphate dehydrogenase deficiency and the need for a novel treatment to prevent kernicterus. Clin Perinatol 43: 341–354.
- 4.↑
Olusanya BO, Emokpae AA, Zamora TG, Slusher TM, 2014. Addressing the burden of neonatal hyperbilirubinaemia in countries with significant glucose-6-phosphate dehydrogenase deficiency. Acta Paediatr 103: 1102–1109.
- 5.↑
Douglas NM, Lampah DA, Kenangalem E, Simpson JA, Poespoprodjo JR, Sugiarto P, Anstey NM, Price RN, 2013. Major burden of severe anemia from non-falciparum malaria species in southern Papua: a hospital-based surveillance study. PLoS Med 10: e1001575; discussion e1001575.
- 6.↑
Kenangalem E, Karyana M, Burdarm L, Yeung S, Simpson JA, Tjitra E, Anstey NM, Poespoprodjo JR, Price RN, Douglas NM, 2016. Plasmodium vivax infection: a major determinant of severe anaemia in infancy. Malar J 15: 321.
- 7.↑
Luxemburger C, van Vugt M, Jonathan S, McGready R, Looareesuwan S, White NJ, Nosten F, 1999. Treatment of vivax malaria on the western border of Thailand. Trans R Soc Trop Med Hyg 93: 433–438.
- 8.↑
Robinson LJ et al. 2015. Strategies for understanding and reducing the Plasmodium vivax and Plasmodium ovale hypnozoite reservoir in Papua New Guinean children: a randomised placebo-controlled trial and mathematical model. PLoS Med 12: e1001891.
- 9.↑
Llanos-Cuentas A et al. 2014. Tafenoquine plus chloroquine for the treatment and relapse prevention of Plasmodium vivax malaria (DETECTIVE): a multicentre, double-blind, randomised, phase 2b dose-selection study. Lancet 383: 1049–1058.
- 10.↑
Baird JK, 2015. Point-of-care G6PD diagnostics for Plasmodium vivax malaria is a clinical and public health urgency. BMC Med 13: 296.
- 11.↑
Ainoon O, Alawiyah A, Yu YH, Cheong SK, Hamidah NH, Boo NY, Zaleha M, 2003. Semiquantitative screening test for G6PD deficiency detects severe deficiency but misses a substantial proportion of partially-deficient females. Southeast Asian J Trop Med Public Health 34: 405–414.
- 12.↑
Espino FE, Bibit JA, Sornillo JB, Tan A, von Seidlein L, Ley B, 2016. Comparison of three screening test kits for G6PD enzyme deficiency: implications for its use in the radical cure of vivax malaria in remote and resource-poor areas in the Philippines. PLoS One 11: e0148172.
- 13.↑
LaRue N et al. 2014. Comparison of quantitative and qualitative tests for glucose-6-phosphate dehydrogenase deficiency. Am J Trop Med Hyg 91: 854–861.
- 14.
Nantakomol D, Paul R, Palasuwan A, Day NP, White NJ, Imwong M, 2013. Evaluation of the phenotypic test and genetic analysis in the detection of glucose-6-phosphate dehydrogenase deficiency. Malar J 12: 289.
- 15.↑
Domingo GJ, Advani N, Satyagraha AW, Sibley CH, Rowley E, Kalnoky M, Cohen J, Parker M, Kelley M, 2018. Addressing the gender-knowledge gap in glucose-6-phosphate dehydrogenase deficiency: challenges and opportunities. Int Health, doi:10.1093/inthealth/ihy060.
- 16.↑
Chu CS et al. 2017. Haemolysis in G6PD heterozygous females treated with primaquine for Plasmodium vivax malaria: a nested cohort in a trial of radical curative regimens. PLoS Med 14: e1002224.
- 17.↑
Bancone G, Gornsawun G, Chu CS, Porn P, Pal S, Bansil P, Domingo GJ, Nosten F, 2018. Validation of the quantitative point-of-care CareStart biosensor for assessment of G6PD activity in venous blood. PLoS One 13: e0196716.
- 18.↑
Bhutani VK, Kaplan M, Glader B, Cotten M, Kleinert J, Pamula V, 2015. Point-of-care quantitative measure of glucose-6-phosphate dehydrogenase enzyme deficiency. Pediatrics 136: e1268–e1275.
- 19.↑
Brito MA, Peixoto HM, Almeida AC, Oliveira MR, Romero GA, Moura-Neto JP, Singh N, Monteiro WM, Lacerda MV, 2016. Validation of the rapid test Carestart(tm) G6PD among malaria vivax-infected subjects in the Brazilian Amazon. Rev Soc Bras Med Trop 49: 446–455.
- 20.↑
Domingo GJ et al. 2013. G6PD testing in support of treatment and elimination of malaria: recommendations for evaluation of G6PD tests. Malar J 12: 391.
- 21.↑
Bancone G, Chu CS, Chowwiwat N, Somsakchaicharoen R, Wilaisrisak P, Charunwatthana P, Bansil P, McGray S, Domingo GJ, Nosten FH, 2015. Suitability of capillary blood for quantitative assessment of G6PD activity and performances of G6PD point-of-care tests. Am J Trop Med Hyg 92: 818–824.
- 22.↑
Kumar R, Indrayan A, 2011. Receiver operating characteristic (ROC) curve for medical researchers. Indian Pediatr 48: 277–287.
- 23.↑
Belfield KD, Tichy EM, 2018. Review and drug therapy implications of glucose-6-phosphate dehydrogenase deficiency. Am J Health Syst Pharm 75: 97–104.
- 24.↑
Howes RE, Battle KE, Satyagraha AW, Baird JK, Hay SI, 2013. G6PD deficiency: global distribution, genetic variants and primaquine therapy. Adv Parasitol 81: 133–201.
- 25.↑
Osorio L et al. 2015. Performance of BinaxNOW G6PD deficiency point-of-care diagnostic in P. vivax-infected subjects. Am J Trop Med Hyg 92: 22–27.
- 26.↑
Rueangweerayut R et al. 2017. Hemolytic potential of tafenoquine in female volunteers heterozygous for glucose-6-phosphate dehydrogenase (G6PD) deficiency (G6PD Mahidol variant) versus G6PD-normal volunteers. Am J Trop Med Hyg 97: 702–711.
- 27.↑
World Health Organization, 2016. Technical Specifications Series for Submission to WHO Prequalification—Diagnostic Assessment: In Vitro Diagnostics Medical Devices to Identify Glucose-6-Phosphate Dehydrogenase (G6PD) Activity. Available at: http://apps.who.int/iris/bitstream/handle/10665/252628/9789241511865-eng.pdf?sequence=1. Accessed October 2, 2018.
- 28.↑
Chiang SH, Fan ML, Hsiao KJ, 2008. External quality assurance programme for newborn screening of glucose-6-phosphate dehydrogenase deficiency. Ann Acad Med Singapore 37: 84–87.
- 29.
Roper DR, De la Salle B, Soni V, Fletcher K, Green JA, 2017. Abrogation of red blood cell G6PD enzyme activity through heat treatment: development of survey material for the UK NEQAS G6PD scheme. Int J Lab Hematol 39: 308–316.
- 30.↑
Tang J, Zhou X, Liu X, Ning L, Zhou W, He Y, 2017. External quality assessment program for detection of glucose-6-phosphate dehydrogenase deficiency in the Guangxi region. Exp Ther Med 14: 2021–2024.
- 31.↑
Tan IK, Whitehead TP, 1969. Automated fluorometric determination of glucose-6-phosphate dehydrogenase (G6PD) and 6-phosphogluconate dehydrogenase (6PGD) activities in red blood cells. Clin Chem 15: 467–478.