Rendell M, Anderson E, Schlueter W, Mailliard J, Honigs D, Rosenthal R, 2003. Determination of hemoglobin levels in the finger using near infrared spectroscopy. Clin Lab Haematol 25 :93–97.
Cheesbrough M, 2000. DHT Haeomoglobinometer. District Laboratory Practice in Tropical Countries. Cambridge: Cambridge University Press, 303–304.
Bentley SA, Johnson A, Bishop CA, 1993. A parallel evaluation of four automated hematology analyzers. Am J Clin Pathol 100 :626–632.
Wardlaw SC, Levine RA, 1983. Quantitative buffy coat analysis. A new laboratory tool functioning as a screening complete blood cell count. JAMA 249 :617–620.
Bland JM, Altman DG, 1986. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1 :307–310.
Past two years | Past Year | Past 30 Days | |
---|---|---|---|
Abstract Views | 260 | 243 | 13 |
Full Text Views | 345 | 4 | 1 |
PDF Downloads | 41 | 5 | 1 |
We have evaluated the accuracy of a simple and inexpensive photometric device (DHT) for the estimation of the blood concentration of hemoglobin by comparison with an automated, high-resolution, flow cytometry-based hematology analyzer (CellDyn 3000) and a centrifugal quantitative buffy coat hematology system (QBC I). We have analyzed the hemoglobin values of 163 individual blood samples. Bland-Altman analysis showed that the methods agreed only poorly: mean differences were 1.0 g/dL with limits of agreement (LOA) of −1.2 g/dL to 3.2 g/dL for the comparison of DHT and CellDyn measurements, 0.5 g/dL with LOA of -2.0 g/dL to 3.0 g/dL for the comparison of DHT with QBC measurements, and 0.5 g/dL with LOA of −1.1 g/dL to 2.1 g/dL for the comparison of QBC with CellDyn measurements. We conclude that the poor agreement of the DHT with the CellDyn does not permit the use of the DHT for critical hemoglobin measurements, particularly in transfusion services.
Rendell M, Anderson E, Schlueter W, Mailliard J, Honigs D, Rosenthal R, 2003. Determination of hemoglobin levels in the finger using near infrared spectroscopy. Clin Lab Haematol 25 :93–97.
Cheesbrough M, 2000. DHT Haeomoglobinometer. District Laboratory Practice in Tropical Countries. Cambridge: Cambridge University Press, 303–304.
Bentley SA, Johnson A, Bishop CA, 1993. A parallel evaluation of four automated hematology analyzers. Am J Clin Pathol 100 :626–632.
Wardlaw SC, Levine RA, 1983. Quantitative buffy coat analysis. A new laboratory tool functioning as a screening complete blood cell count. JAMA 249 :617–620.
Bland JM, Altman DG, 1986. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1 :307–310.
Past two years | Past Year | Past 30 Days | |
---|---|---|---|
Abstract Views | 260 | 243 | 13 |
Full Text Views | 345 | 4 | 1 |
PDF Downloads | 41 | 5 | 1 |