Abstract

Abstract.

Most malaria testing is by rapid diagnostic tests (RDTs) that detect histidine-rich protein 2 (HRP2). Recently, several RDT manufacturers have developed highly sensitive RDTs (hsRDTs), promising a limit of detection (LOD) orders of magnitude lower than conventional RDTs. To model the added utility of hsRDTs, HRP2 concentration in Angolan outpatients was measured quantitatively using an ultrasensitive bead-based assay. The distribution of HRP2 concentration was bimodal in both afebrile and febrile patients. The conventional RDT was able to detect 81% of all HRP2-positive febrile patients and 52–77% of HRP2-positive afebrile patients. The added utility of hsRDTs was estimated to be greater in afebrile patients, where an hsRDT with a LOD of 200 pg/mL would detect an additional 50–60% of HRP2-positive persons compared with a conventional RDT with a LOD of 3,000 pg/mL. In febrile patients, the hsRDT would detect an additional 10–20% of cases. Conventional RDTs already capture the vast majority of symptomatic HRP2-positive individuals, and hsRDTs would have to reach a sufficiently low LOD approaching 200 pg/mL to provide added utility in identifying HRP2-positive, asymptomatic individuals.

Loading

Article metrics loading...

/content/journals/10.4269/ajtmh.17-0262
2017-10-11
2017-11-21
Loading full text...

Full text loading...

http://instance.metastore.ingenta.com/content/journals/10.4269/ajtmh.17-0262
Loading
/content/journals/10.4269/ajtmh.17-0262
Loading

Data & Media loading...

Supplementary Data

Supplemental Figure

  • Received : 30 Mar 2017
  • Accepted : 07 May 2017

Most Cited This Month

This is a required field
Please enter a valid email address
Approval was a Success
Invalid data
An Error Occurred
Approval was partially successful, following selected items could not be processed due to error